![Tettex TTR 2795 Operating Instructions Manual Download Page 39](http://html1.mh-extra.com/html/tettex/ttr-2795/ttr-2795_operating-instructions-manual_1092947039.webp)
Setting Up Tests
35
7
Setting Up Tests
7.1
Setting Up Tests
To set-up a new test on the system the set-up menu should be selected from the startup display. Press “
UP
” until the
left most button indicates “
SETUP
”. Press this button and the system will enter the setup menu. The following screen
is shown:
Basically the TTR2795 has 3 operating modes; Manual, Semi-automatic and Full-automatic.
Manual
User has to set the HV Config, LV Config and Ph Displacement
Semi-automatic
User has to set the HV Config & LV Config. Ph Displacement is set as ??
“??” selection means that the TTR will detect the Phase Displacement
Full-automatic
User sets
HV Config
to “
Auto 3 ph
”. “
Auto 3 ph
” selection means that the
TTR will try to detect the vector group of the transformer
HV Config
The
HV Config
sets the winding configuration of the HV side of the transformer.
The options for setting the “
LV Config
” and “
Ph Displacement
” may or may not
be displayed depending on the HV configuration selected.
Config
LV Config
Ph Displ.
Comments
Auto 3 Ph
Hidden
Hidden
Attempt to automatically determine
config
D
Shown
Shown
Delta (D) Winding
Y
Shown
Shown
Star (Wye) Winding
Yn
Shown
Shown
Star (Wye) Winding with Neutral
Z
Shown
Shown
Zig-Zag Winding
Zn
Shown
Shown
Zig-Zag Winding with Neutral
CT
Hidden
Hidden
Current Transformer
Single
Hidden
Hidden
Single Phase
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...