Remote Control
67
<IB> (float) is the measured energization current for phase B
<PB> (float) is the measured phase shift for phase B
<TRC> (float) is the measured turns ratio for phase C
<IC> (float) is the measured energization current for phase C
<PC> (float) is the measured phase shift for phase C
<Pass> (integer) indicates whether all three phases are within the acceptable
percentage
deviation from nominal. If <Pass> is zero, one or more of the ratios are out of
specification.
If it is non zero they are within acceptable limits.
Comments:
None
Test:Results:Leg
This returns the results of the currently measured leg(s) (Leg A, Leg B, Leg C).
If Leg A was measured, it returns Leg A results and Leg B, C results are zero.
If Leg B was measured, it returns Leg A, B results and Leg C results are zero.
If Leg C was measured, it returns Leg A, B, C results.
Structure:
+T:R:L:<TapNum>:~:
Fields:
<TapNum> (integer) is the number of the tap for which the results are to be shown.
If the
transformer is not tapped, this should be 0. The maximum value is <NumTaps>
Returns:
+OK:<HV>:<LV>:<TRA>:<IA>:<PA>:<TRB>:<IB>:<PB>:<TRC>:
<IC>:<PC>:<Pass>:~:
- The results of the test
+ERROR:0907:~:
- The value of <TapNum> is out of range
+ERROR:090E:~:
- The value of <NumTap> is valid, but that tap has not
been
measured. This will occur if the test was halted before completion
<HV> (float) is the boilerplate HV side voltage for this tap
<LV> (float) is the boilerplate LV side voltage for this tap
<TRA> (float) is the measured turns ratio for phase A, if Leg A was measured,
otherwise 0.
<IA> (float) is the measured energization current for phase A, if Leg A was
measured,
otherwise 0.
<PA> (float) is the measured phase shift for phase A, if Leg A was measured,
otherwise 0.
<TRB> (float) is the measured turns ratio for phase B, if Leg B was measured,
otherwise 0.
<IB> (float) is the measured energization current for phase B, if Leg B was
measured,
otherwise 0.
<PB> (float) is the measured phase shift for phase B, if Leg B was measured,
otherwise 0.
<TRC> (float) is the measured turns ratio for phase C, if Leg C was measured,
otherwise 0.
Summary of Contents for TTR 2795
Page 1: ...Operating Instructions HAEFELY TEST AG TTR 2795 Transformer Turns Ratio Meter Version 1 4 ...
Page 10: ...6 Introduction Maximum Roll Diameter Outer 50mm Inner 10mm ...
Page 22: ...18 Connection and Setup Figure 11 Connecting to three phase BCT ...
Page 101: ...Application Software APSW 2795 97 ...
Page 111: ...Miscellaneous 107 14 7 Conformity ...