Configure the Instrument
Configure and Select Signal Inputs
87
PRISM MPI-MPX User Manual
4.
Select
IP
.
5.
Select
ST2022-6
.
6.
Slide the ST2022-7 Seamless Switching
control to Enabled or Disabled.
Note:
There is a difference between “port” and “path”. The port is the physical SFP
port that is used to input the 10/25GE signal. Path 1 and Path 2 are the signal paths to
be used for seamless switching. This differentiation is being made because it is
possible for both Path 1 and Path 2 to use a single port.
Note:
With ST2022-7 seamless switching, a reconstructed output stream is created
from the two input streams. The reconstructed stream is used downstream for various
display tiles like Picture. The buffer size to absorb the packet reception time difference
between Path1 and Path2 is 10ms. The output stream reconstruction does not work
properly for streams on Path1 and Path2 with skew greater than 10ms. This error
makes tiles show the wrong images and information.
7.
Enter the Source and Destination information for Path 1 and Path 2, if applicable.
Note:
Path 2 is only available if ST2022-7 Seamless Switching is enabled. You may
have to scroll down to see Path 2.
8.
To have the instrument ignore a parameter, select the
Unmasked
control next to
the parameter to change the state to Masked.
Summary of Contents for PRISM MPI
Page 1: ...PRISM MPI and MPX SDI IP Waveform Monitor User Manual April 2021 D00010020C...
Page 6: ...6 PRISM MPI MPX User Manual...
Page 10: ...10 PRISM MPI2 25 MPX2 25 User Manual...
Page 54: ...Getting Started Activity Center PRISM MPI MPX User Manual 54...
Page 76: ...Methods of Operation Remote Control through API Commands PRISM MPI MPX User Manual 76...
Page 138: ...Configure the Instrument Troubleshooting PRISM MPI MPX User Manual 138...