Set and Recall Instrument Presets
Instrument setups can become complex as you tailor them to monitor various
parameters. To save set up time and ensure consistency, you can save and quickly
access instrument setups in your instrument using the
Presets
icon (
) on the
instrument display.
Note:
For the MPX monitor, you can recall instrument presets from the front panel.
(See
.)
The instrument can store up to 36 separate instrument presets. The presets are divided
into six groups, A through F, with each group containing 6 preset storage locations. The
default preset names are A1-A6, B1-B6, C1-C6, D1-D6, E1-E6, and F1-F6.
All instrument settings except those set in the Utilities and Network Settings submenus
of the Configuration menu are saved in a Preset. Settings that are not saved are those
such as the clock setting, IP configuration, and network address. This is an overview of
the settings saved in a Preset. Some settings are option dependent.
•
Application assigned to each tile
•
Application icon order in the application selector list
•
Application specific settings
•
Input selection and configuration
•
PTP reference settings
•
SDI Generator and IP Generator settings
Turn on Edit Mode
When Edit mode is on, selecting a preset button displays a menu to allow changes to
that preset button (Rename and Save as) you can also recall the existing settings
(Recall).
When edit mode is off, selecting a preset button only recalls the preset settings
assigned to that button, with no confirmation.
Summary of Contents for PRISM MPI
Page 1: ...PRISM MPI and MPX SDI IP Waveform Monitor User Manual April 2021 D00010020C...
Page 6: ...6 PRISM MPI MPX User Manual...
Page 10: ...10 PRISM MPI2 25 MPX2 25 User Manual...
Page 54: ...Getting Started Activity Center PRISM MPI MPX User Manual 54...
Page 76: ...Methods of Operation Remote Control through API Commands PRISM MPI MPX User Manual 76...
Page 138: ...Configure the Instrument Troubleshooting PRISM MPI MPX User Manual 138...