Configure the Instrument
Configure Reference Settings
PRISM MPI-MPX User Manual
112
2.
Select
Reference
to open the Reference submenu and select the
Analog Ref
Setting
submenu.
3.
Select from one of these reference I/O options:
–
Loop-Through:
The REF input is looped out the REF output BNC.
–
Terminate Input:
The REF input is terminated at the REF input BNC. There is no
signal at REF output BNC.
Configure the PTP Reference Settings
1.
Select the
Settings
icon (
) to open the Settings menu.
2.
Select
Reference
to open the Reference submenu and select the
PTP Settings
submenu.
3.
To enable or disable PTP logging to the Event Log, select the control box to change
the setting to the state.
4.
Select the profile to configure: ST2059, AES67, or General. The default
communication mode for each profile is Multicast.