Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 271
TRIGger:A:HOLdoff:VALue
Sets or queries the A trigger holdoff time as a percent of record. This is
equivalent to setting
Holdoff (% of record)
in the
Mode & Holdoff
side menu.
Trigger
TRIGger:A:HOLdoff:VALue <NR3>
TRIGger:A:HOLdoff:VALue?
<NR3>
from 0 to 100 and is a percent of the holdoff range.
TRIGGER:A:HOLDOFF:VALUE 50
to set the holdoff value to 50% of record.
TRIGger:A:LEVel
Sets the A trigger level. This command is equivalent to adjusting the front-panel
Trigger
LEVEL
knob.
Trigger
TRIGger:A:LEVel { ECL | TTL | <NR3> }
TRIGger:A:LEVel?
ECL
specifies a preset ECL level of --1.3 V.
TTL
specifies a preset TTL level of 1.4 V.
<NR3>
is the A trigger level, in volts.
TRIGGER:A:LEVEL?
might return
:TRIGGER:A:LEVEL 1.4E0
, indicating that the A edge trigger level
is set to 1.4 V.
TRIGGER:A:LEVEL 0.5
sets the A trigger level to 0.5 V.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...