Command Descriptions
2- 306
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
TRIGger:STATE? (Query Only)
Returns the current state of the triggering system.
Trigger
TRIGger:STATE?
AUTO
indicates that the oscilloscope is in auto mode and acquires data even in the
absence of a trigger.
ARMED
indicates that the oscilloscope is acquiring pretrigger information. All
triggers are ignored when TRIGger:STATE is ARMed.
READY
indicates that all pretrigger information has been acquired and the
oscilloscope is ready to accept a trigger.
SAVe
indicates that acquisition is stopped or that all channels are off.
TRIGger
indicates that the oscilloscope has seen a trigger and is acquiring the
posttrigger information.
TRIGGER:STATE?
might return
ARMED
, indicating that pretrigger data is being acquired.
Group
Syntax
Returns
Examples