Command Descriptions
2- 70
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
CH<x>:VOLts
Sets or queries the vertical gain of the specified channel. This command is
identical to the CH<x>:SCAle command and is included for compatibility
purposes. Only CH<x>:SCAle is returned in response to a CH<x>? query.
Vertical
CH<x>:VOLts <NR3>
CH<x>:VOLts?
<x>
specifies the input channel. Valid values are 1, 2, 3, and 4.
<NR3>
is the gain, in user units per division. The range is 10 V per division to
1 mV per division when using a 1x, 1 megohm, voltage probe.
CH4:VOLTS 2.00E0
sets the channel 4 gain to 2.00 units per division.
CH2:VOLTS?
might return
1.0E+0
, indicating that the current units per division setting of
channel 2 is 1 unit per division.
CH<x>:SCAle
CH<x>:YUNit
Sets or returns the units of the specified channel.
Vertical
CH<x>:YUNit <QString>
CH<x>:YUNit?