Command Descriptions
2- 210
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
<x>
specifies which waveform to use, where <x> = 1 is the “from” waveform,
and <x> = 2 is the “to” waveform.
FALL
specifies the falling edge.
RISe
specifies the rising edge.
MEASUREMENT:IMMED:DELAY:EDGE1 RISE
specifies that the “from” waveform rising edge be used for the immediate delay
measurement.
MEASUREMENT:IMMED:DELAY:EDGE1?
returns either
RISE
or
FALL
.
MEASUrement:IMMed:SOUrce
Specifies the source to measure “from” when taking an immediate delay
measurement or phase measurement. This command is the same as the MEA-
SUrement:IMMed:SOURCE1 command.
MEASUrement:IMMed:SOURCE<x>
Specifies the source to measure “from” or “to” when taking an immediate delay
measurement or phase measurement.
Measurement
MEASUrement:IMMed:SOURCE<x> { CH<x> | MATH | MATH1 | REF<x> }
MEASUrement:IMMed:SOURCE?
SOURCE<x>
specifies which waveform to use, where <x> = 1 is the “from”
waveform, and <x> = 2 is the “to” waveform.
CH<x>
is an input channel.
MATH
or
MATH1
is the math waveform.
REF<x>
is a reference waveform.
MEASUREMENT:IMMED:SOURCE1 MATH1
specifies MATH1 as the immediate measurement source.
Arguments
Examples
Group
Syntax
Arguments
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...