Command Descriptions
2- 226
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:MEAS<x>:VALue? (Query Only)
Returns the value that has been calculated for the measurement specified by <x>.
NOTE
. This value is a display value and is updated on the oscilloscope screen
approximately every 1/3 second.
Measurement
MEASUrement:MEAS<x>:VALue?
MEASUREMENT:MEAS1:VALUE?
might return
MEASUREMENT:MEAS1:VALUE 3.742E-6
, indicating the measure-
ment value displayed at the time the command was executed.
MEASUrement:MEAS<x>:DATa?
MEASUrement:METHod
Sets or returns the method used to calculate the 0% and 100% reference level.
This is equivalent to setting the
High-Low Setup
in the
MEASURE
menu.
Measurement
MEASUrement:METHod { Auto | HIStogram | MINMax }
MEASUrement:METHod?
HIStogram
sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax
sets the high and low waveform levels to MAX and MIN, respectively.
Auto
chooses the best method (HIStogram or MINMax) for each dataset.
Group
Syntax
Examples
Related Commands
Group
Syntax
Arguments
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...