![Tektronix TDS3000 Series Program Manual Download Page 294](http://html.mh-extra.com/html/tektronix/tds3000-series/tds3000-series_program-manual_1079160294.webp)
Command Descriptions
2- 256
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
*SAV (Set Only)
(Save) stores the state of the oscilloscope into a specified internal setup location.
You can later use the *RCL command to restore the oscilloscope to this saved
state. This is equivalent to selecting the
Save Current Setup
menu item in the
SAVE/RECALL
menu.
Save and Recall
*SAV <NR1>
<NR1>
is a value in the range from 1 to 10 and specifies a location. Any settings
that have been stored previously at this location will be overwritten.
*SAV 2
saves the current settings in internal setup location 2.
FACtory, *RCL, RECAll:SETUp, SAVe:SETUp
SAVe:MASK
(Requires TDS3TMT)
Saves the User mask to the specified location on the disk or flash drive.
Save and Recall
SAVe:MASK <QString>
<QString>
is the path and file name, enclosed in quotes, to which to save the
User mask data on the disk or flash drive.
SAVE:SETUP “COAX2.MSK“
saves the user mask information to the file
COAX2.MSK
in the current directory.
Group
Syntax
Arguments
Examples
Related Commands
Group
Syntax
Arguments
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...