Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 57
*CAL? (Query Only)
Runs an internal self-calibration and returns the oscilloscope calibration status.
NOTE
. Disconnect or otherwise removeall input signals prior to starting
self-calibration.
The self-calibration can take several minutes to complete. No other commands
are executed until calibration is complete.
Calibration and Diagnostic
*CAL?
<NR1>
= 0 indicates that the calibration did not complete successfully.
<NR1>
¸
0 indicates that the calibration completed without errors.
*CAL?
performs an internal self-calibration and might return
1
to indicate that the
calibration was successful.
CALibrate:FACtory
This command is the same as CALibrate:FACtory:STARt.
CALibrate:FACtory:ABOrt (Set Only)
Causes the oscilloscope to abort the factory calibration process, reset to the
calibration initialization step, and exit without saving calibration data or status
information.
Calibration and Diagnostic
CALibrate:FACtory:ABOrt
CALIBRATE:FACTORY:ABORT
aborts the current factory calibration process.
Group
Syntax
Returns
Examples
Group
Syntax
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...