Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 173
OFF
turns off mask lock to waveform.
ON
turns on mask lock to waveform.
<NR1>
is an integer number where NR1 = 0 turns off mask lock to waveform,
and NR1
≠
0 turns on mask lock to waveform.
MASK:LOCk OFF
turns off mask lock to waveform.
MASK:MARGIN:VERTical
(Requires TDS3TMT)
Sets or returns the User mask’s vertical margin percentage. This command
redraws the vertical mask margins by the specified percentage, where the vertical
margin percent is the percentage of the User mask’s nominal pulse amplitude.
This command affects all segments in the User mask. Segment vertical margins
move away from the nominal waveform pulse position for positive percentage
values, and towards the nominal waveform pulse position for negative percent-
age values. This command only operates on User (custom) masks.
Mask
MASK:MARGIN:VERTical <NR3>
MASK:MARGIN:VERTical?
<NR3>
is a real number that sets the User mask margins as a percentage of the
user mask’s nominal pulse amplitude.
MASK:MARGIN:VERTICAL 5
Increases the User mask vertical dimensions by 5%.
Arguments
Examples
Group
Syntax
Arguments
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...