Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 77
CURSOR:HBARS:SELECT CURSOR1
selects the first horizontal bar cursor as the active cursor.
CURSOR:HBARS:SELECT?
returns
:CURSOR:HBARS:SELECT CURSOR1
when the first cursor is the active
cursor.
CURSor:HBArs:UNIts
Sets or queries the measurement units for the horizontal bar cursors.
Cursor
CURSor:HBArs:UNIts <QString>
CURSor:HBArs:UNIts?
<QString> is V for volts, A for amps, AA for amps
2
, A/V for amps/volts, V/A
for volts/amps, VV for volts
2
, W for watts, % for percentage, or IRE when the
IRE graticule is selected.
CURSOR:HBARS:UNITS?
returns
:CURSOR:HBARS:UNITS ”IRE”
when the horizontal bar cursor units are
IRE.
CURSor:HBArs:USE (Set Only)
Sets the horizontal bar cursor measurement scale. This command is only
applicable when ratio cursors are on.
Cursor
CURSor:HBArs:USE { CURrent | FIVEdivs }
<CURrent>
sets the H Bar measurement scale so that 0% is the current position
of the lowest H Bar cursor and 100% is the current position of the highest H Bar
cursor. Moving the cursors after executing this command does not change the 0%
or 100% scale points.
Examples
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...