Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 63
CALIBRATE:INTERNAL:STATUS?
might return
:CALIBRATE:INTERNAL:STATUS UNCAL+
if non-volatile memory
has failed.
CALibrate:INTERNal:STARt
CH<x>? (Query Only)
Returns the vertical parameters of the specified channel. Because CH<x>:SCAle
and CH<x>:VOLts are identical, only CH<x>:SCAle is returned.
Vertical
CH<x>?
CH1?
might return the string
:CH1:SCALE 10.0E-3;POSITION 0.0E0;
OFFSET 0.0E0;COUPLING DC;BANDWIDTH FULL;DESKEW 0.0E0;
IMPEDANCE MEG;PROBE 1.0E0;YUNIT ”V”;INVERT 0
for channel 1.
CH<x>:BANdwidth
Sets or queries the bandwidth setting of the specified channel. This is equivalent
to setting
Bandwidth
in the
Vertical
menu.
Vertical
CH<x>:BANdwidth { TWEnty | ONEFifty | FULl }
CH<x>:BANdwidth?
<x>
specifies the input channel. Valid values are 1, 2, 3, and 4.
TWEnty
sets the channel bandwidth to 20 MHz.
ONEFifty
sets the channel bandwidth to 150 MHz. This argument is not
available on some models.
FULl
sets the channel bandwidth to the full bandwidth of the oscilloscope.
Examples
Related Commands
Group
Syntax
Examples
Group
Syntax
Arguments
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...