Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 157
LIMit:HARDCopy { OFF | ON | <NR1> }
LIMit:HARDCopy?
ON
or
<NR1>
¸
0 turns on the hard copy operation when any waveform data
exceeds the limits set by the limit test.
OFF
or
<NR1>
= 0 turns off the hard copy operation.
LIMit:HARDCopy ON
specifies that the hard copy operation occurs when any waveform data exceeds
the limits specified in the limit test.
LIMit:HARDCopy?
returns either
0
or
1
, indicating whether the hard copy operation occurs when any
waveform data exceeds the limits specified in the limit test.
LIMit:COMPARE:CH<x>, LIMit:STATE, HARDCopy
LIMit:SAVEWFM
(Requires TDS3LIM)
Sets or returns the TDS3LIM save waveform to file on limit test failure setting.
When enabled, this command copies waveform data from all active waveforms
that have limit tests failures to a file on the disk or flash drive. The file format is
specified by the SAVe:WAVEform:FILEFormat command. Set the file save
format to Internal File Format to minimize file size.
Limit Test
LIMit:SAVEWFM { OFF | ON | <NR1> }
LIMit:SAVEWFM?
OFF
or
<NR1>
= 0 turns off save waveform to disk or flash drive for limit testing.
ON
or
<NR1>
¸
0 turns on save waveform to disk or flash drive for limit testing.
LIMIT:SAVEWFM ON
specifies that save waveform to disk or flash drive is in effect.
Syntax
Arguments
Examples
Related Commands
Group
Syntax
Arguments
Examples
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...