![Tektronix TDS3000 Series Program Manual Download Page 133](http://html.mh-extra.com/html/tektronix/tds3000-series/tds3000-series_program-manual_1079160133.webp)
Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 95
DATA:SOURCE REF2
specifies that reference waveform two will be transferred in the next CURVE?
query.
DATA:SOURCE?
might return
CH1
, indicating that channel 1 is the source for the waveform data
that is transferred using a CURVE? query.
DATa:STARt
Sets or returns the starting data point for waveform transfer. This command
allows for the transfer of partial waveforms to and from the oscilloscope.
Waveform
DATa:STARt <NR1>
DATa:STARt?
<NR1>
ranges from 1 to the record length, and is the first data point that will be
transferred. Data will be transferred from
<NR1>
to DATa:STOP or the record
length, whichever is less. If
<NR1>
is greater than the record length then no data
will be transferred.
When DATa:STOP is less than DATa:STARt, the values are swapped internally
for CURVe?
DATA:START 10
specifies that the waveform transfer will begin with data point 10.
DATA:START?
might return
214
as the first waveform data point that will be transferred.
CURVe?, DATa:STOP
Examples
Group
Syntax
Arguments
Examples
Related Commands
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...