Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 319
Y
specifies a normal waveform where one ASCII or binary data point is
transmitted for each point in the waveform record. Only
y
values are explicitly
transmitted. Absolute coordinates are given by:
X
n
=
XZEro
+
XINcr
(
n–PT_Off
)
Y
n
=
YZEro
+
YMUlt
(
y
n
−
YOFf
)
ENV
specifies that the oscilloscope transmit the waveform as minimum and
maximum point pairs. Peak detect and envelope waveforms use an ENV format.
Only y values are explicitly transmitted. Absolute coordinates are given by:
X
n
=
XZEro
+
XINcr
(
n–PT_Off
)
Y
nmin
=
YZEro
+
YMUlt
(
y
nmin
−
YOFf
)
Y
nmax
=
YZEro
+
YMUlt
(
y
nmax
−
YOFf
)
WFMPRE:PT_FMT Y
sets the waveform data point format to Y.
WFMPRE:PT_FMT?
returns either
Y
or
ENV
.
WFMPre:PT_Off
The set form of this command is ignored. The query form always returns a 0, if
the waveform specified by DATA:SOUrce is on or displayed. If the waveform is
not displayed, the query form generates an error and returns event code 2244.
This command is listed for compatibility with other Tektronix oscilloscopes.
Waveform
WFMPre:PT_Off
WFMPre:PT_Off?
Arguments are ignored.
WFMPre:<wfm>:PT_Off
Arguments
Examples
Group
Syntax
Arguments
Related Commands
Summary of Contents for TDS3000 Series
Page 8: ...Table of Contents vi TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 10: ...Preface viii TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 38: ...Getting Started 1 28 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 82: ...Command Groups 2 44 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...
Page 388: ...Status and Events 3 20 TDS3000 TDS3000B and TDS3000C Series Programmer Manual ...