Replacement Procedures
6–8
RFM 150 Instruction Manual
7. Replace speaker or speaker gasket (optional):
a. Remove the speaker and gasket.
b. Install the replacement gasket. Position the speaker on the
gasket.
8. Replace the keypads (optional):
a. Lift out the two keypads.
b. Install the replacement keypads in the front instrument casing,
using a dental pick or similar instrument to press the keypad
under the speaker housing.
9. Place the replacement Front Panel circuit board assembly in
position. Place the stiffener on top of the Front Panel circuit
board assembly. Install and partially tighten the 15 Front Panel
circuit board assembly screws. From the front of the instrument,
check the alignment of the keypads by pressing firmly on each
key. Make sure the keys are making contact and adjust if
necessary. Tighten all Front Panel circuit board assembly screws,
beginning with those around the speaker, to 4 in-lbs.
10. Solder the speaker and encoder switch leads to the replacement
Front Panel circuit board set, as shown in Figure 6–3.
11. Clean the clear filter on the instrument front case with a soft,
lint-free cloth dampened with a non-abrasive liquid glass cleaner.
12. Perform Instrument Assembly on page 6–15.
13. Perform the following steps from the Performance Check,
beginning on page 5–6:
Step 1, Step 2, and Steps 9 through 20.
Replacement Procedures
6–8
RFM 150 Instruction Manual
7. Replace speaker or speaker gasket (optional):
a. Remove the speaker and gasket.
b. Install the replacement gasket. Position the speaker on the
gasket.
8. Replace the keypads (optional):
a. Lift out the two keypads.
b. Install the replacement keypads in the front instrument casing,
using a dental pick or similar instrument to press the keypad
under the speaker housing.
9. Place the replacement Front Panel circuit board assembly in
position. Place the stiffener on top of the Front Panel circuit
board assembly. Install and partially tighten the 15 Front Panel
circuit board assembly screws. From the front of the instrument,
check the alignment of the keypads by pressing firmly on each
key. Make sure the keys are making contact and adjust if
necessary. Tighten all Front Panel circuit board assembly screws,
beginning with those around the speaker, to 4 in-lbs.
10. Solder the speaker and encoder switch leads to the replacement
Front Panel circuit board set, as shown in Figure 6–3.
11. Clean the clear filter on the instrument front case with a soft,
lint-free cloth dampened with a non-abrasive liquid glass cleaner.
12. Perform Instrument Assembly on page 6–15.
13. Perform the following steps from the Performance Check,
beginning on page 5–6:
Step 1, Step 2, and Steps 9 through 20.
Summary of Contents for RFM150
Page 4: ......
Page 14: ...Table of Contents x RFM150 Instruction Manual Table of Contents x RFM150 Instruction Manual...
Page 21: ...Getting Started Getting Started...
Page 22: ......
Page 45: ...Operating Basics Operating Basics...
Page 46: ......
Page 93: ...Reference Reference...
Page 94: ......
Page 123: ...Specifications Specifications...
Page 124: ......
Page 134: ...Specifications 4 10 RFM150 Instruction Manual Specifications 4 10 RFM150 Instruction Manual...
Page 136: ......
Page 137: ...Performance Check Performance Check...
Page 138: ......
Page 151: ...Maintenance Maintenance...
Page 152: ......
Page 177: ...Replaceable Electrical Parts Replaceable Electrical Parts...
Page 178: ......
Page 185: ...Replaceable Mechanical Parts Replaceable Mechanical Parts...
Page 186: ......
Page 197: ...Appendices Appendices...
Page 198: ......
Page 269: ...Index Index...
Page 270: ......
Page 277: ......
Page 278: ......