Performance Check
RFM 150 Instruction Manual
5–3
Required Equipment
RF Signal Source
Frequency Range: 10 MHz to 1000 MHz
Accuracy:
±
500 Hz
Amplitude Range: 0 dBm maximum output (50
W
or 75
W
)
FM Modulation (internal or external): Deviation range 5 kHz to
25 kHz. Rate between 20 Hz and 7.5 kHz.
AM Modulation (internal or external): Amplitude range 0 to 2.5%
(corresponds to 0 to 5% RFM 150 reading). Rate 50, 60, 100 or
120 Hz.
Example: Hewlett-Packard 8657A Signal Generator with a Tektronix
CFG250 Function Generator
The function generator is used when testing hum accuracy, to drive
the external modulation input to generate the 50, 60, 100, or 120 Hz
AM modulation rate.
Function Generator
For testing hum accuracy, use a function generator to drive the
external modulation input of the RF signal source to generate a
60 Hz sine wave.
Example: Tektronix CFG250 Function Generator
Power Meter With
75
W
Sensor Head
Sensitivity: –20 dBm
Frequency Range: 10 MHz to 1 GHz
Accuracy:
±
0.1 dB
Example: Hewlett-Packard 438A Power Meter with Hewlett-Packard
8483A Sensor Head
Performance Check
RFM 150 Instruction Manual
5–3
Required Equipment
RF Signal Source
Frequency Range: 10 MHz to 1000 MHz
Accuracy:
±
500 Hz
Amplitude Range: 0 dBm maximum output (50
W
or 75
W
)
FM Modulation (internal or external): Deviation range 5 kHz to
25 kHz. Rate between 20 Hz and 7.5 kHz.
AM Modulation (internal or external): Amplitude range 0 to 2.5%
(corresponds to 0 to 5% RFM 150 reading). Rate 50, 60, 100 or
120 Hz.
Example: Hewlett-Packard 8657A Signal Generator with a Tektronix
CFG250 Function Generator
The function generator is used when testing hum accuracy, to drive
the external modulation input to generate the 50, 60, 100, or 120 Hz
AM modulation rate.
Function Generator
For testing hum accuracy, use a function generator to drive the
external modulation input of the RF signal source to generate a
60 Hz sine wave.
Example: Tektronix CFG250 Function Generator
Power Meter With
75
W
Sensor Head
Sensitivity: –20 dBm
Frequency Range: 10 MHz to 1 GHz
Accuracy:
±
0.1 dB
Example: Hewlett-Packard 438A Power Meter with Hewlett-Packard
8483A Sensor Head
Summary of Contents for RFM150
Page 4: ......
Page 14: ...Table of Contents x RFM150 Instruction Manual Table of Contents x RFM150 Instruction Manual...
Page 21: ...Getting Started Getting Started...
Page 22: ......
Page 45: ...Operating Basics Operating Basics...
Page 46: ......
Page 93: ...Reference Reference...
Page 94: ......
Page 123: ...Specifications Specifications...
Page 124: ......
Page 134: ...Specifications 4 10 RFM150 Instruction Manual Specifications 4 10 RFM150 Instruction Manual...
Page 136: ......
Page 137: ...Performance Check Performance Check...
Page 138: ......
Page 151: ...Maintenance Maintenance...
Page 152: ......
Page 177: ...Replaceable Electrical Parts Replaceable Electrical Parts...
Page 178: ......
Page 185: ...Replaceable Mechanical Parts Replaceable Mechanical Parts...
Page 186: ......
Page 197: ...Appendices Appendices...
Page 198: ......
Page 269: ...Index Index...
Page 270: ......
Page 277: ......
Page 278: ......