Performance Check
5–6
RFM 150 Instruction Manual
Performance Check Procedure
Record the results of this procedure on your copy of Table 5–1.
1. Verify battery and POWER key operation as follows:
a. With the AC adapter disconnected from the instrument, press
[POWER] to turn on the RFM 150.
b. A low-pitched beep followed by a high-pitched beep will
signal that the instrument is initializing. The instrument title
screen will be displayed briefly, followed by a measurement
screen.
c. Press [POWER] to turn off the instrument. Check that the
screen is blank.
2. Verify operation from an external power source as follows:
a. With the instrument turned off, plug the AC adapter into the
DC input connector.
b. Plug the AC adapter into a compatible power source.
c. The instrument will emit two short beeps, and the screen will
remain blank.
d. Press [POWER] to turn on the RFM 150.
e. A low-pitched beep followed by a high-pitched beep will
signal that the instrument is initializing. The instrument title
screen will be displayed briefly, followed by a measurement
screen.
3. Use the following key sequence to make the CATV–STD channel
table the active table:
[UTIL] [F2] {use knob to select CATV–STD} [F1] [ESC].
4. Use the following key sequence to select the 1 channel
measurement mode:
[LEVEL] [F1].
5. Verify amplitude accuracy
a. Use the following key sequence to tune the RFM 150 to
10 MHz: [1] [0] [MHz].
Performance Check
5–6
RFM 150 Instruction Manual
Performance Check Procedure
Record the results of this procedure on your copy of Table 5–1.
1. Verify battery and POWER key operation as follows:
a. With the AC adapter disconnected from the instrument, press
[POWER] to turn on the RFM 150.
b. A low-pitched beep followed by a high-pitched beep will
signal that the instrument is initializing. The instrument title
screen will be displayed briefly, followed by a measurement
screen.
c. Press [POWER] to turn off the instrument. Check that the
screen is blank.
2. Verify operation from an external power source as follows:
a. With the instrument turned off, plug the AC adapter into the
DC input connector.
b. Plug the AC adapter into a compatible power source.
c. The instrument will emit two short beeps, and the screen will
remain blank.
d. Press [POWER] to turn on the RFM 150.
e. A low-pitched beep followed by a high-pitched beep will
signal that the instrument is initializing. The instrument title
screen will be displayed briefly, followed by a measurement
screen.
3. Use the following key sequence to make the CATV–STD channel
table the active table:
[UTIL] [F2] {use knob to select CATV–STD} [F1] [ESC].
4. Use the following key sequence to select the 1 channel
measurement mode:
[LEVEL] [F1].
5. Verify amplitude accuracy
a. Use the following key sequence to tune the RFM 150 to
10 MHz: [1] [0] [MHz].
Summary of Contents for RFM150
Page 4: ......
Page 14: ...Table of Contents x RFM150 Instruction Manual Table of Contents x RFM150 Instruction Manual...
Page 21: ...Getting Started Getting Started...
Page 22: ......
Page 45: ...Operating Basics Operating Basics...
Page 46: ......
Page 93: ...Reference Reference...
Page 94: ......
Page 123: ...Specifications Specifications...
Page 124: ......
Page 134: ...Specifications 4 10 RFM150 Instruction Manual Specifications 4 10 RFM150 Instruction Manual...
Page 136: ......
Page 137: ...Performance Check Performance Check...
Page 138: ......
Page 151: ...Maintenance Maintenance...
Page 152: ......
Page 177: ...Replaceable Electrical Parts Replaceable Electrical Parts...
Page 178: ......
Page 185: ...Replaceable Mechanical Parts Replaceable Mechanical Parts...
Page 186: ......
Page 197: ...Appendices Appendices...
Page 198: ......
Page 269: ...Index Index...
Page 270: ......
Page 277: ......
Page 278: ......