Removal and Installation Procedures
6–62
CSA8000 & TDS8000 Instruments and Sampling Modules
Exchanging electrical sampling module
If it is determined that a sampling module failure is caused by an internal part,
you must exchange the complete sampling module chassis to Tektronix. The
customer will keep the cover and attaching hardware. Sampling module
exchange can be made with either your local Tektronix service center or central
Tektronix exchange in Beaverton, Oregon.
1. Transport the chassis sampling modules with terminations on the inputs.
2. Transport the chassis sampling modules in their original containers and store
the sampling modules on a metal or conductive surface.
3. When handling the exposed chassis sampling module only, do not touch any
exposed component on the circuit board. Grasp the exposed module by the
edges of the circuit board or the rear or front housings.
Program the serial number (printed on the sampling module cover) into the new
replacement sampling module chassis.
1. Reinstall the chassis into the cover following the procedure on page 6–61.
2. Install the electrical sampling module into one of the instrument’s slots.
3. Connect a GPIB cable to the GPIB port of the instrument connect the
remaining end to the GPIB port of a PC.
CAUTION.
Do not have any user interface dialogs active. Dialogs such as
Compensation, System properties will prevent the instrument from responding to
GPIB commands.
4. Enter the following GPIB commands:
a. SYSTem:PROTect OFF
b. SYSTem:PROPerties:CH(1-8):SERialnum <serialNum_string>
Example: SYST:PROP:CH3:SER “B010123”
NOTE
. For a 2 channel sampling module either channel can be specified in
step b. For an 80E01 sampling module use only the lower numbered channel
number such as CH1 or CH3.
c.
SYSTem:PROTect ON
Replacing the sampling
module chassis
Reinstalling exchange
module serial number
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......