CSA8000 & TDS8000 Instruments and Sampling Modules
4–3
Brief Procedures
There are three procedures in this section that provide a quick way to confirm
basic functionality and proper adjustment:
H
Perform the Diagnostics
H
Perform the Compensation
H
Perform the Functional Tests
These procedures require little or no test equipment; see the description on
page 4–1 for information on when to use these procedures.
Perform the Diagnostics
Do the following steps to run the internal routines that confirm basic functional-
ity and proper adjustment.
Equipment required
None
Prerequisites
First, all sampling modules to be diagnosed must be installed as
outlined in their user manuals.
Second, power on the instrument and allow a 20 minute warm-up
before doing this procedure.
1. Set up the instrument: From the application menu bar, select Utilities, and
then select Diagnostics. The Diagnostics dialog box displays. See below.
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......