Electrical Sampling Modules
4–62
CSA8000 & TDS8000 Instruments and Sampling Modules
Random noise, displayed
These procedures check those characteristics that relate to the signal acquisition
and are listed as checked under Warranted Characteristics in Specifications.
Refer to Table 1–12, page 1–12.
Equipment
required
No external equipment required
Prerequisites
As listed on page 4–19 and under Prerequisites on page 4–53.
No test equipment hook up is required.
1. Set up the CSA8000/TDS8000 as follows:
a. Select C1 from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
b. In the Trig Setup dialog box, set Trigger Source to Internal Clock (200
KHz).
c.
In the Acq Setup dialog box:
H
set Acquisition Mode to Sample.
H
under Stop After, check Condition and select Number of Acquisi-
tions from the pull down menu. Enter 96 for the number of
acquisitions.
d. In the Horz Setup dialog box:
H
set Record length to 1000 points.
H
set Scale to 1 ps/division.
H
set Position to 0 (this will revert to a minium non-zero value).
H
set Horizontal Reference to 0%.
e.
In the Vert Setup dialog box:
H
set Scale to 2 mV/division if testing with an 80E01 module; set to
1 mV/division for all other modules.
H
set Offset to 0 mV.
Follow this procedure to make the Random noise checks:
1. WfmDatabase menu, click Display Database, select C1
2. Histogram menu, select vertical, Enable Histogram.
Setup
Test
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......