Adjustment Procedures
5–2
CSA8000 & TDS8000 Instruments and Sampling Modules
Required Equipment
The adjustment procedure requires the specific test equipment and materials
listed in Table 5–2.
Table 5–2: Required equipment and materials
Category
Specific equipment required
Quantity
Signal source
Signal generator 50 MHz to 1 GHz,
1 ppm frequency
accuracy
1 ea
Meter
Digital Multimeter, with 6.5 plus digits
1 ea
Instrument
controller
(only one of
PC-compatible computer with National Instruments
GPIB Controller card and software, running Windows
95/98 or NT
1 ea
these configura-
tions required)
National Instruments USB GPIB Controller card for
Windows 98 and National Instruments NI-488.2 for
Windows software (to install in the instrument)
1
1 ea
Divider
Power, 50
W
, SMA “T”. male Tektronix 015-0565-00
1 ea
Adapter
SMA “T”, male to 2 SMA female Tektronix 015-1016-00
1 ea
Adapter
SMA male to BNC female, Tektronix 015-0554-00
1 ea
Adapter
BNC to dual banana plug, Tektronix 103-0095-00
1 ea
Coaxial cables
50
W
, male-to-male SMA connectors
3 ea
GPIB cables
GPIB cable, 1 m minimum length, Tektronix 002-0991-01
1 ea
1
Depending on other USB devices attached to the instrument, a USB hub may also be
required. Consult your National Instruments documentation for more information.
Instrumentation Setup
Set the National Instruments GPIB Interface command software to allow GPIB
communications between the PC controller and CSA8000/TDS8000, device
under test (DUT).
Before you can execute the adjustment program, you need to set up and
configure the instrumentation. Make connections according to the following
diagrams. See Figure 5–1.
To perform this procedure you must have the specific equipment listed in
Table 5–2.
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......