Main Instrument
CSA8000 & TDS8000 Instruments and Sampling Modules
4–43
External direct trigger delay jitter, short term optimized and locked to internal
10 MHz reference modes
These procedures check those characteristics that relate to the system trigger and
are listed as checked under Warranted Characteristics in Specifications. Refer to
Table 1–3, page 1–3.
Equipment
required
Synthesized signal generator (Item 13)
0.5 meter precision SMA cable (Item 10)
1.0 meter precision SMA cable (Item 11)
Power divider (Item 9)
80E0X sampling module
Prerequisites
As listed on page 4–19 and under Prerequisites on page 4–34.
Install the test hookup:
1. Install the 80E0X module into the left-most small module compartment of
the TDS8000 or CSA8000 instrument (see Figure 4–13 on page 4–43).
2. Connect the synthesized signal generator to a 1.0 meter cable.
3. Connect the remaining end of the 1.0 meter cable to a SMA power divider.
Connect two 0.5 meter cables to the two remaining SMA power divider
connectors.
4. Connect one of the 0.5 meter cables to Channel 1 of the 80E0X module.
Connect the remaining 0.5 meter cable to the TRIGGER DIRECT INPUT.
CSA8000/TDS8000
Synthesized signal generator
Channel 1
0.5 Meter precision cables
Power divider
TRIGGER
DIRECT INPUT
1.0 Meter precision cable
RF Output
Figure 4–13: External direct trigger jitter test hookup
Setup
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......