80C00 Modules Specifications
1–32
CSA8000 & TDS8000 Instruments and Sampling Modules
Table 1–19: Optical modules – Clock recovery option (CR)
Name
Characteristics
Effective wavelength
Module
Range
range
80C01
1270 nm to 1600 nm
80C02
1270 nm to 1600 nm
80C03
700 nm to 1650 nm
80C04
1270 nm to 1600 nm
Operating data rates
80C01:
622.08 MHz
±
1000 ppm (OC-12/STM-4)
2.48832 GHz
±
1000 ppm (OC-48/STM-16)
80C02:
9.95328 Gb/s
"
1000 ppm (OC-192/STM-64)
80C03:
1.0625 Gb/s
"
1000 ppm (FC1063)
1.2500 Gb/s
"
1000 ppm (GBE)
2.48832 Gb/s
"
1000 ppm (OC-48/STM-16)
2.5000 Gb/s
"
1000 ppm (2X GBE)
80C04-CR1:
9.95328 Gb/s
"
1000 ppm (OC-192/STM-64)
80C04-CR2:
9.95328 Gb/s
"
1000 ppm (OC-192/STM-64),
10.664 Gb/s
"
1000 ppm (FEC)
The incoming data stream must be of non-return-to-zero format (NRZ)
and, for OC-192/STM-64, must have a data sequence content which
provides both isolated 1s and multi-consecutive mark sequences (that
is 2,3,4 and so forth logical 1s in a consecutive row). NOTE: a fixed
pattern of 10101010. . . does not meet that requirement; the 80C02-CR
clock recovery function may not properly lock to such a pattern. The
80C02-CR will, however, typically lock to a 11001100. . . pattern (this
is equivalent to a 2.48832 GHz optical square wave).
n
Optical sensitivity
range, clock recover
(optical input power)
1
Module Maximum
Minimum
80C01
+ 5.0 dBm (3.16 mW), typical
–10.0 dBm (100
m
W), typical
80C02
+7 dBm (5.0 mW), typical
–10.0 dBm (100
m
W), typical
–7.5 dBm, warranted
80C03
–4.0 dBm (400
m
W), warranted –16.0 dBm (25
m
W),
warranted
80C04
+7 dBm (5.0 mW), typical
–10.0 dBm (100
m
W), typical
–7.5 dBm, warranted
Clock and data
electrical output
amplitudes
2
80C01:
u
300 mV
pp,
typical
80C02
Serial DATA output:
u
700 mV
pp,
typical
Serial CLOCK output:
1.5 V
pp,
typical
1/16th CLOCK output:
600 mV
pp,
typical
80C03:
u
350 mV
pp,
typical
80C04
Serial DATA output:
u
700 mV
pp,
typical
Serial CLOCK output:
1.5 V
pp,
typical
1/16th CLOCK output:
600 mV
pp,
typical
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......