d. In the Horz Setup dialog box:
H
set Record Length to 2000 points.
H
set Position to 0 (this will revert to a minimum non-zero value).
H
set Horizontal Reference to 0%.
e.
In the Vert Setup dialog box:
H
set Scale to 35 mV/division.
H
set Offset to 0.
Follow this procedure for the 80E01 sampling module reference signals:
1. In the Meas Setup dialog box:
a. make sure the channel under test is selected in the Source menu.
b. select Meas 1, and then choose Meas > Amplitude > Amplitude.
2. Set the RF signal frequency listed in the Frequency column, Table 4–6 on
page 4–71.
3. Set the RF signal level listed in the Reference Set Level dBm column,
Table 4–6.
4. In the Horz Setup dialog box, set the CSA8000/TDS8000 Horizontal Scale
to (1/Frequency), where Frequency is the RF signal frequency.
5. Press the CLEAR DATA button on the front panel of the
CSA8000/TDS8000.
6. Press the RUN/STOP button on the front panel of the CSA8000/TDS8000.
7. Record amplitude reading readout and use the value as AMPLITUDE (Volts)
in the step that follows.
8. Convert Volts to dBm as follows:
Vp=(AMPLITUDE/2)
Vrms=Vp/SQRT (2)
P1=Vrms*Vrms/50
dBm=10*log(P1/0.001)
9. Record dBm in the 80E01 Measured level column, Table 4–6 on page 4–71.
10. Repeat steps 2 through 9 for all frequencies listed in the Frequency column,
Table 4–6 on page 4–71.
Tests
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......