Adjustment Procedures
5–4
CSA8000 & TDS8000 Instruments and Sampling Modules
CALIBRATE:DCCALIBRATOR 0.0
Record DMM reading.
7. Enter the following GPIB command to set the instrument’s DC calibrator
offset cal constant to –1 * DMM reading:
CALCOMP:DOUBLE “DcCalOffsetAdj”,(–1.0 * the DMM reading)
Example: CALCOMP:DOUBLE “DcCalOffsetAdj”, 3.2e–4
8. Enter the following GPIB command to set the instrument’s
calibrator to 1.0 V:
CALIBRATE:DCCALIBRATOR 1.0
Record the DMM reading (reading1).
9. Enter the following GPIB command to set the instrument’s
calibrator to –1.0 V:
CALIBRATE:DCCALIBRATOR –1.0
Record the DMM reading (reading2).
Calculate cal constant as follows:
(–1.0 * Reading2
)
Reading1)
2
10. Enter the following GPIB command to set the instrument’s
calibrator Lsb cal constant:
CALIBRATE:DOUBLE “DcCalLsbAdj”,calculated result from Step 9.
11. Wait 8 seconds then enter the following GPIB command to save the DC
adjustments as follows:
CAL:SAVE:FACT:MAI
12. Enter the following GPIB command to turn the instrument’s cal constant
protection on:
SYST:PROT ON
To verify that the DC calibration adjustment was successful complete the
following procedures:
1. Set the instrument’s DC calibrator to –1.0 V and verify as follows:
CALIBRATE:DCCALIBRATOR –1.0
2. Verify that the DMM reads –1.0 V
"
1 mV
3. Set the instrument’s DC calibrator to 1.0 V and verify as follows:
CALIBRATE:DCCALIBRATOR 1.0
4. Verify that the DMM reads 1.0 V
"
1 mV
5. Set the instrument’s DC calibrator to 0.0 V and verify as follows:
CALIBRATE:DCCALIBRATOR 0.0
DC Calibrator Adjust
Verification
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......