Removal and Installation Procedures
6–48
CSA8000 & TDS8000 Instruments and Sampling Modules
1. Locate module to be removed: Locate the Acquisition assembly in the locator
diagram, Figure 6–9, page 6–20. Remove the Acquisition assembly
following in this order:
H
Trim
H
Bottom cover
H
Left and Right covers
H
Front and Rear Power Distribution Boards
2. Orient the instrument: Set the instrument so its top is down on the work
surface and its bottom is facing you.
3. Remove the Acquisition assembly: The Acquisition assembly includes, the
Acquisition board and Electrical Front End board. See Figure 6–29,
page 6–50.
a. Remove the
5
/
16
inch nut that secures the TRIGGER PRESCALE SMA
cable to the front panel.
b. Remove the nine T-15 Torxdrive screws securing the Acquisition
assembly to the chassis.
NOTE
. Six Acquisition assembly screws are 0.312 inch long. Three Acquisition
assembly screws are 1.750 inch long. The 1.750 inch screw locations are marked
on the circuit board.
c.
Disconnect the two SMB connectors that are connected to the trigger
hybrid, located on the front edge of the Acquisition circuit board.
d. Disconnect two coaxial cables from J8 and J16, located on the Acquisi-
tion circuit board.
e.
Disconnect J10 ribbon cable connector.
f.
Orient the instrument: Set the instrument so its front panel surface is
facing you. Push the module slot covers in to access the slot connectors
and attaching hardware.
CAUTION.
Note the locations of the T-10 Torxdrive screws and the
3
/
16
inch hex
threaded posts for later reassembly. Make sure you reinstall the Torxdrive screw
and threaded posts in the correct position. The
3
/
16
inch hex threaded posts have
etched lines around their mounting holes. Interchanging the threaded posts with
the Torxdrive screws could cause damage to your module connector. See refer to
Figure 6–28 for hardware locations.
Acquisition Assembly
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......