Installation
2–6
CSA8000 & TDS8000 Instruments and Sampling Modules
Table 2–2: Line fuses
Fuse type
Rating
Fuse part number
Cap & fuseholder
part number
0.25 x 1.250 inch
8 A, fast blow, 250 V
159-0046-00
200-2264-00
5 x 20 mm
6.3 A, fast blow, 250 V
159-0381-00
200-2265-00
Fuses
AC Power
Power Switch
Figure 2–4: Line fuse and power cord connector locations, rear panel
CAUTION.
Connect the keyboard, mouse, and other accessories before applying
power to the product. Connecting the accessories after powering on the
instrument can damage the accessories. Two exceptions are the USB keyboard
and mouse that ships with the instrument. Both can be plugged or unplugged
without first turning power off.
2. Connect the keyboard and mouse, observing the caution above. Note that the
instrument ships with a USB keyboard, which plugs into the USB port (see
Figure 2–3 on page 2–5 for location) and a USB mouse, which plugs into
the back of the USB keyboard.
NOTE
. Connection of the keyboard and mouse is optional. You can operate most
features without them, using the front-panel controls and the touchscreen.
3. Connect the power cord.
4. If you have an external monitor, connect the power cord and power on the
monitor.
5. Turn the Power switch on at the rear panel. (See Figure 2–4 on page 2–6 for
switch location.)
6. Push the On/Standby switch to power on the instrument (see Figure 2–5 for
the switch location).
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......