Electrical Sampling Modules
CSA8000 & TDS8000 Instruments and Sampling Modules
4–67
Analog bandwidth (80E01)
These procedures check those characteristics that relate to the signal acquisition
and are listed as checked under Warranted Characteristics in Specifications.
Refer to Table 1–12, page 1–12. This check has two procedures that characterizes
the reference signals. The last procedure characterizes the 80E01 sampling
module.
Equipment
required
Synthesized signal generator (Item 13)
Power meter (Item 14)
Power sensor (Item 15)
Adapter, 2.4mm male-to-male (Item 16)
Adapter, 2.4mm female-to-female (Item 17)
Attenuator,10 dB pad (Item 20)
Cable, special, power/phase stability (Item 21)
Prerequisites
As listed on page 4–19 and under Prerequisites on page 4–53.
Install the test hookup to measure the power reference:
NOTE
. Use a torque wrench when connecting the terminators, adapters,
attenuator, and power sensor in this setup. Use 8
"
0.3 inch pounds of torque.
1. Connect the 10 dB attenuator to the Synthesized signal generator. Connect
the remaining end to the power sensor. Connect the power sensor to the
power meter. See Figure 4–19 on page 4–67.
Power meter
Power Sensor
w/attached cable
10dB Attenuator
Synthesized signal generator
RF Output
Figure 4–19: Power-reference hookup
Power-Reference
Characterization Setup
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......