Brief Procedures
4–12
CSA8000 & TDS8000 Instruments and Sampling Modules
After verifying the channels, you can now verify that the time bases function.
This verification is done using a front-panel signal.
Equipment re-
quired
One SMA cable, such as Tektronix part number 174-1427-00.
One 10x SMA attenuator, such as Tektronix 015-1003-00
One electrical (80E00-series) sampling module
Prerequisites
None
1. Initialize the instrument: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the SMA cable from the Internal Clock
output through a 10x attenuator to any 80E00 sampling module input
channel as shown in Figure 4–6.
CSA8000/TDS8000
SMA cable from
INTERNAL CLOCK
output to 80E00 C3 input
10x attenuator
Figure 4–6: Hookup for the time base tests
3. Set up the instrument:
a. Push the Trigger MENU front-panel button to display the Trig Setup
dialog box.
b. Click Internal Clock under Trigger Source in the Trig Setup dialog box.
The Internal Clock rate should be set to 200kHz.
c.
Push the Trigger MENU front-panel button again to dismiss the Trig
Setup dialog box.
d. Push the channel button for the channel you connected to in step 2. The
button lights and the channel display comes on. See Figure 4–7,
page 4–13.
e.
Turn the Vertical SCALE knob to set the vertical scale to 20 mV/div.
The channel scale readout is displayed in the Control bar at the bottom
of the graticule.
Verify the
Time Bases Work
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......