Main Instrument
CSA8000 & TDS8000 Instruments and Sampling Modules
4–41
6. Set the synthesized output to 400 mV peak-to-peak (200 mV displayed)
7. Make the following settings to the CSA8000/TDS8000:
a. Select Default Setup from the Utilities menu on the menu bar at top of
display.
b. Select C1 from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
c.
In the Trig Setup dialog box:
H
select External Direct.
H
select Metastablility Reject.
H
select Auto Mode.
d. In the Vert Setup dialog box:
H
set Scale to 20 mV/division.
H
set Offset to 0.
e.
In the Horz Setup dialog box:
H
set Scale to 200 ps/division.
H
set Horizontal Reference to 0%.
H
set Position to 19 ns.
Follow this procedure to make the direct trigger sensitivity check:
1. In the Meas Setup dialog box:
a. make sure the channel under test is selected in the Source menu.
b. select Meas 1, and then choose Select Meas > Amplitude > Amplitude.
c.
select Meas 2, and then choose Select Meas > Timing > Frequency.
2. Adjust the synthesizer signal generator level until the amplitude measure-
ment readout on the instrument screen indicates the amplitude is 100 mV.
3. Set the signal generator frequency to 100 MHz.
4. Adjust the Horizontal Scale to allow at least two full periods to be displayed.
5. Verify that the frequency read out on the instrument screen is within 10%
from the true value.
Test:
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......