Theory of Operation
3–8
CSA8000 & TDS8000 Instruments and Sampling Modules
hardware filter designed to result in the combined system having an
GBE (1.25 Gb/s Gigabit Ethernet) Reference Receiver response.
For major functional circuit blocks refer to (Figure 9–9 on page 9–9).
The 80C03 module can be optionally configured with clock recovery that
supports Fibre Channel 1063 (1.063 Gbps) and OC-48 / STM-16 (2.488 Gpbs)
standards.
The 80C04 module is optimized for testing of long-wavelength (1100–1650 nm)
signals at either 9.953 Gb/s or 10.664 Gb/s. With its high optical bandwidth
>28 GHz, it is also well suited to general-purpose, high–performance optical
component testing.
H
OC192 or 10.664 Gb/s: The electrical sampler is adjusted to give the
desired response, and signal is not routed through any filter (signal is
sent through the straight-through path). The sampler bandwidth and
response is optimized during the calibration such that the combined
system will have an OC192 (STM-64) or 10.66 Gb Reference Receiver
response.
H
20 GHz: The sampler bandwidth and response is set during the
calibration such that the combined system will have an Optical
Bandwidth (–6dB) > 20 GHz.
H
30 GHz: The sampler bandwidth and response is set during the
calibration such that the combined system will have an Optical
Bandwidth (–6dB) > 28 GHz.
The 80C04 can be optionally configured with clock recovery (Opt. CR-1) that
supports 9.953 Gb/s telecom standards.
For major functional circuit blocks refer to (Figure 9–11 on page 9–11).
The 80C05 module is designed to test long wavelength (1520–1580nm) signals.
This module is intended to be used as a test and measurement tool for high
bandwidth telecommunications with its high optical bandwidth >40 GHz.
H
OC192: There is one Reference Receiver setup selectable for 9.95328
Gb/s SONET/SDH standard.
H
There is no clock recovery option available.
For major functional circuit blocks refer to (Figure 9–13 on page 9–13).
80C04 Optical Sampling
Module
80C05 Optical Sampling
Module
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...