Maintenance
6–4
CSA8000 & TDS8000 Instruments and Sampling Modules
Clean the flat panel display surface by gently rubbing the display with a
clean-room wipe (such as Wypall Medium Duty Wipes, #05701, available from
Kimberly-Clark Corporation).
If the display is very dirty, moisten the wipe with distilled water or a 75%
isopropyl alcohol solution and gently rub the display surface. Avoid using excess
force or you may damage the plastic display surface.
CAUTION.
To prevent getting moisture inside the instrument during external
cleaning, use only enough liquid to dampen the cloth or applicator.
Inspection — Interior.
To access the inside of the instrument for inspection and
cleaning, refer to the Removal and Installation Procedures in this section.
Inspect the internal portions of the instrument for damage and wear, using Table
6–2 as a guide. Defects found should be repaired immediately.
If any circuit board is replaced, check Table 5–1 on page 5–1, in Section 5 to see
if it is necessary to adjust the instrument.
CAUTION.
To prevent damage from electrical arcing, ensure that circuit boards
and components are dry before applying power to the instrument.
Table 6–2: Internal inspection check list
Item
Inspect for
Repair action
Circuit boards
Loose, broken, or corroded
solder connections. Burned
circuit boards. Burned, broken, or
cracked circuit-run plating.
Remove and replace damaged
circuit board.
Resistors
Burned, cracked, broken, blis-
tered condition.
Remove and replace damaged
circuit board.
Solder connections
Cold solder or rosin joints.
Resolder joint and clean with
isopropyl alcohol.
Capacitors
Damaged or leaking cases.
Corroded solder on leads or
terminals.
Remove and replace damaged
circuit board.
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......