Brief Procedures
4–6
CSA8000 & TDS8000 Instruments and Sampling Modules
b. Wait until the Status for all items you wish to compensate changes from
Warm Up to Pass, Fail, or Comp Req’d.
c.
Under Select Action, click the Compensate option button.
d. From the top pulldown list, choose All (default selection) to select the
main instrument and all its modules as targets to compensate.
e.
Click the Execute button to begin the compensation.
f.
Follow the instructions to disconnect inputs and install terminations that
will appear on screen; be sure to follow static precautions (see the user
manual for your sampling module) when following these instructions.
NOTE
. Failing to install the 50 ohm terminations on 80E00 sampling-module
input connectors can yield erroneous compensation failures or results.
2. Verify that the compensation routines pass:
a. The compensation may take several minutes to complete. Verify that
Pass appears as Status for the main instrument and for all sampling
modules listed in the Compensation dialog box when compensation
completes.
b. If instead Fail appears as Status, rerun the compensation. If Fail status
continues after rerunning compensation and you have allowed warm up
to occur, the module or main instrument may need service.
c.
If you want to save the compensation constants generated by this
compensation, click the Save option button under Select Action. Click
the Execute button to save the compensation.
End of Procedure
Summary of Contents for CSA8000 Series
Page 4: ......
Page 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 23: ......
Page 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Page 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 64: ......
Page 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 93: ......
Page 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 104: ......
Page 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 219: ......
Page 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 308: ......
Page 313: ......
Page 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Page 316: ......