Calibration—
7D14
CHANNEL A
CHECK
Equipment
Required
1.
Indicator
Oscilloscope
7.
GR 50-ohm
in-line termination
2.
Amplifier unit
8.
GR to
BNC male
adapter
3. Time-base unit
9. 5-nanosecond
GR cable
4.
High-frequency constant-amplitude sine-wave generator
10. 42-inch
BNC cable
5.
Medium-frequency constant-amplitude
sine-wave gen-
11. 18-inch BNC cable
erator
12.
BNC T connector
6.
Low-frequency constant-amplitude sine-wave generator
13. GR 10X attenuator
Control
Settings
Set
the
controls as given under
Preliminary Control
Settings.
NOTE
In steps one through four,
the
exact reading obtained
depends
on
the accuracy of the generator dial setting.
1.
Check
CH A INPUT Frequency Range
a.
Connect
the high-frequency
constant-amplitude
sine
wave generator to the amplifier unit input through the five-
nanosecond
GR cable, 10X GR attenuator, and
50-ohm
in-line termination, in given
order.
b.
Set
the amplifier unit for a vertical deflection factor
of 50
millivolts/division.
c.
Set the
high-frequency generator for a two-division
(100
mV)
display at
the 3-MHz reference frequency.
d.
Disconnect
the 50-ohm in-line termination from the
amplifier
unit
input,
and reconnect
it to the CH A INPUT
connector.
e. Set
the high-frequency generator for a
500-MHz out
put.
f.
CHECK—The
readout
display
should show a steady
count
of
about 500.00000 MHz. Change the dial setting if
necessary to obtain a count of at least 500 MHz.
NOTE
This procedure
checks the
7D14 frequency range to
500
MHz.
To check the frequency range to the maxi
mum
limit as given under Performance
Requirement
in
Section
1, use a constant-amplitude generator with
at least a 525-MHz
output.
g.
Set the INPUT SENS switch to 50 Ω/100
mV.
h.
Replace
the 50-ohm in-line termination on the
high-
frequency
generator output
with a GR to BNC male
adapter, and
reconnect the generator to the CH A INPUT
connector.
i.
CHECK—
The
readout
display for
the same count
obtained
in part f of this step (at least 500 MHz).
j.
Disconnect
all test equipment.
2.
Check
CH
A
INPUT
50
Ω Sensitivity
and
Low-
Frequency
Range
(AC-COUPLING)
a.
Change the following control
settings:
7D14
INPUT SENS
50Ω/1V
MEASUREMENT
INTERVAL
1s
b. Connect the low-frequency constant-amplitude gener
ator to
the amplifier unit input connector through the 42-
inch
BNC cable and BNC
T connector. Connect the output
5-5
Summary of Contents for 7D14
Page 4: ...7D14 ...
Page 11: ...Operating Instructions 7D14 Fig 2 1 7D14 front panel controls and connectors 2 2 ...
Page 33: ... 3 13 Fig 3 11 Logic diagram for Zero Cancel Logic stage Circuit Description 7D14 ...
Page 38: ...3 18 Fig 3 16 Time Base and Control circuit detailed block diagram Circuit Description 7D14 ...
Page 44: ...NOTES ...
Page 46: ...NJ Fig 4 1 Electrode configuration for semiconductors in this instrument I ...
Page 68: ...NOTES ...
Page 96: ... 7DI 4 DIGITAL COUNTER UNIT ...
Page 98: ...GRS 0371 BLOCK DIAGRAM ...
Page 99: ......
Page 103: ...0 0 I 200 mV 500 µs 0 001 200 mV 500 µs 0 001 MHz 00 mV E 00 µs 0 001 MHz ...
Page 106: ...1 ...
Page 110: ...A2 Logic Circuit Board Assembly jQ798i 798 jc743 CR744t uni R724 R742 JL744S FrR796 ...
Page 113: ......
Page 114: ...P 0 A2 LOGIC BOARD ...
Page 121: ... A B D t F H J NPR Tj ZWX V A AC M AEA HUM qAZ DEF HJ N P R S T J V V7X y ABM APAI AA 7DI4 ...
Page 129: ......
Page 130: ...4 7D14 DIGITAL COUNTER ...