Calibration—
7D14
of
the
BNC
T connector to the CH A INPUT through the
18-inch
BNC cable.
c. Set the amplifier
unit for a vertical deflection factor
of
.2 volts/division.
g.
Press the
COUPLING switch AC button.
h.
CHECK
—Readout display for a steady count at
about
5 Hz (0.005
kHz).
d.
Set
the low-frequency generator for a five-division
display
(1 volt) at 200 kHz.
e.
CHECK
—Readout
display for a steady count at
about
200
kHz.
f.
Press the
COUPLING switch AC button.
g. CH
ECK—Readout display for the
same
count
obtained
in part e of this
step.
h.
Set the
low-frequency generator for a five-division
display (1
volt) at 10 kHz.
i.
CHECK—Readout display for
zero count
(0.000 to
0.001
kHz).
3.
Check
CH A
INPUT
1 MΩ
Sensitivity
and
Low-
Frequency
Range
a.
Change
the following control settings:
7D14
INPUT SENS
1
MΩ/1 V
COUPLING
DC
b.
CHECK
—Readout
display for
a steady count at about
10 kHz.
c.
Set
the
INPUT SENS switch to 1 MΩ/10 V and
the
amplifier
unit for a vertical deflection factor of 2 volts/
division.
d. Set
the low-frequency generator for a five-division
display (10 volts)
at 10 kHz.
e. CHECK—Readout display for a steady count at
about
10 kHz.
4.
Check
TRIG
SOURCE
Sensitivity
a.
Change
the following control settings:
7D14
INPUT SENS
TRIG
SOURCE
COUPLING
DC
b.
Set the low-frequency
generator
for a 1
,5-division dis
play
(3 volts) at 10 kHz.
c.
CHECK
—Readout
display for a steady count at about
10 kHz.
d. Disconnect
all
test equipment.
5.
Check
BW-5
MHz
a.
Install
the 7D14 in the left vertical plug-in compart
ment of the Indicator Oscilloscope.
b.
Change
the following control settings:
Indicator
Oscilloscope
Vertical
Mode
Chop
7D14
INPUT SENS
50Ω/100mV
BW
5
MHz
c.
Connect
the medium-frequency constant-amplitude
generator
to
the CH A INPUT connector through the five-
nanosecond GR cable, GR to BNC
male adapter, and
BNC
T
connector. Connect the output of the BNC T
connector
to
the amplifier unit input through
the 18-inch BNC cable.
f.
Set
the low-frequency generator for a five-division
dis
play (10 volts) at 5
Hz.
d. Set the
amplifier
unit for a vertical deflection factor
of
20 millivolts/division.
5-6
@i
Summary of Contents for 7D14
Page 4: ...7D14 ...
Page 11: ...Operating Instructions 7D14 Fig 2 1 7D14 front panel controls and connectors 2 2 ...
Page 33: ... 3 13 Fig 3 11 Logic diagram for Zero Cancel Logic stage Circuit Description 7D14 ...
Page 38: ...3 18 Fig 3 16 Time Base and Control circuit detailed block diagram Circuit Description 7D14 ...
Page 44: ...NOTES ...
Page 46: ...NJ Fig 4 1 Electrode configuration for semiconductors in this instrument I ...
Page 68: ...NOTES ...
Page 96: ... 7DI 4 DIGITAL COUNTER UNIT ...
Page 98: ...GRS 0371 BLOCK DIAGRAM ...
Page 99: ......
Page 103: ...0 0 I 200 mV 500 µs 0 001 200 mV 500 µs 0 001 MHz 00 mV E 00 µs 0 001 MHz ...
Page 106: ...1 ...
Page 110: ...A2 Logic Circuit Board Assembly jQ798i 798 jc743 CR744t uni R724 R742 JL744S FrR796 ...
Page 113: ......
Page 114: ...P 0 A2 LOGIC BOARD ...
Page 121: ... A B D t F H J NPR Tj ZWX V A AC M AEA HUM qAZ DEF HJ N P R S T J V V7X y ABM APAI AA 7DI4 ...
Page 129: ......
Page 130: ...4 7D14 DIGITAL COUNTER ...