7D14
SECTION
S
CALIBRATION
Change
information, if
any, affecting this section will be found
at the rear of this manual.
Recalibration Interval
To
assure instrument accuracy, check the calibration of
the
7D14 every 1000 hours
of operation, or every six
months
if
used
infrequently.
Before complete calibration,
thoroughly
clean and inspect this instrument as outlined in
the Maintenance section.
Tektronix
Field
Service
Tektronix, Inc.
provides complete instrument repair and
recalibration
at local Field Service Centers and the Factory
Service
Center. Contact your
local Tektronix Field Office
or representative
for further information.
Using This
Procedure
General.
This
section provides several features to facili
tate
checking
or adjusting the 7D14. These are:
Index.
To aid
in
locating a
step in the Performance
Check
or Adjustment procedure, an index is given pre
ceding Part
I — Performance Check
and
Part II — Adjust
ment
procedure.
Performance
Check.
The
performance
of this instru
ment can
be checked
without removing the side shields or
making
internal adjustments by performing only Part I —
Performance
Check. This procedure checks the
instrument
against
the tolerances listed in the Performance Require
ment
column of Section 1. In addition, a cross-reference is
provided
to the step in Part II — Adjustment, which will
return
the
instrument to correct calibration. In most cases,
the adjustment
step can be performed without
changing
control settings
or equipment connections.
Adjustment
Procedure.
To
return this instrument
to
correct
calibration with the minimum number of steps, per
form
only Part II — Adjustment. The Adjustment proce
dure
gives the recommended calibration procedure
for all
circuits
in this instrument.
Partial
Procedure.
A
partial check or
adjustment
is often
desirable
after replacing
components
or to touch up the
adjustment
of a portion of
the instrument
between major
recalibrations. To
check or adjust only
part of the instru
ment, start
with the nearest
Equipment Required list pre
ceding
the desired
portion.
To prevent unnecessary recali
bration
of other parts
of the instrument, readjust only if
the
tolerance given in the CHECK—
part of the step is not
met.
If re adjustment is necessary, also check the calibra
tion
of any
steps listed in the INTERACTION— part of the
step.
Complete Performance
Check
/Adjustment. To
completely
check
and adjust all parts of this instrument,
perform
both
Parts I and II. Start the complete procedure
by
performing the Adjustment procedure for a portion
of
the
instrument (e.g., Channel
A Adjustments) and follow
this
with
the Performance Check for the same portion (e.g.,
Channel
A Checks).
This method will assure that the instru
ment is
both
correctly adjusted and performing within all
Performance
Requirements
as
given
in Section 1.
TEST
EQUIPMENT
REQUIRED
General
The
following test
equipment and accessories, or its
equivalent,
is
required for
complete calibration of the
7D14.
Specifications
given for
the test equipment are the
minimum
necessary
for
accurate
calibration. Therefore,
some
of
the specifications
listed here may
be somewhat less
precise than
the actual performance capabilities of the test
equipment.
All
test equipment is assumed
to be correctly
calibrated and operating within the listed specifications.
The Performance Check
and Adjustment procedures are
based
on this recommended equipment. If other equipment
is substituted,
control
settings or
calibration
setup may
need
to be altered to meet
the
requirements of the equip
ment
used.
Detailed operating instructions for the test
equipment
are not given in
this procedure. Refer to the
instruction
manual
for the test equipment
if more infor
mation
is needed.
Calibration
Equipment
Alternatives
All
of
the test equipment is required to completely
check
and adjust this
instrument. However, some of
the
items used only for
the
Performance Check
can be deleted
without compromising
the
instrument's measurement
capabilities.
For example, the pulse generator
is used only
in the
Performance Check and
may be deleted if the user
does not
desire to
check the performance of the External
Gate
or External Reset functions.
Equipment used only for
®î
5-1
Summary of Contents for 7D14
Page 4: ...7D14 ...
Page 11: ...Operating Instructions 7D14 Fig 2 1 7D14 front panel controls and connectors 2 2 ...
Page 33: ... 3 13 Fig 3 11 Logic diagram for Zero Cancel Logic stage Circuit Description 7D14 ...
Page 38: ...3 18 Fig 3 16 Time Base and Control circuit detailed block diagram Circuit Description 7D14 ...
Page 44: ...NOTES ...
Page 46: ...NJ Fig 4 1 Electrode configuration for semiconductors in this instrument I ...
Page 68: ...NOTES ...
Page 96: ... 7DI 4 DIGITAL COUNTER UNIT ...
Page 98: ...GRS 0371 BLOCK DIAGRAM ...
Page 99: ......
Page 103: ...0 0 I 200 mV 500 µs 0 001 200 mV 500 µs 0 001 MHz 00 mV E 00 µs 0 001 MHz ...
Page 106: ...1 ...
Page 110: ...A2 Logic Circuit Board Assembly jQ798i 798 jc743 CR744t uni R724 R742 JL744S FrR796 ...
Page 113: ......
Page 114: ...P 0 A2 LOGIC BOARD ...
Page 121: ... A B D t F H J NPR Tj ZWX V A AC M AEA HUM qAZ DEF HJ N P R S T J V V7X y ABM APAI AA 7DI4 ...
Page 129: ......
Page 130: ...4 7D14 DIGITAL COUNTER ...