Chapter 5 Multi-Instrument Operation
Multiple Phase Operation (continued)
Multiple Phase Operation (continued)
Case 3 --
Trigger-Sequential
This is a special case of TRIG mode which allows you to output the signal from each unit sequentially.
Both units are in TRIG mode. Upon receiving the trigger signal (external or manual), the master unit
starts to run while the slave unit waits to receive the SYNC TRIG pulse from the master unit. Upon
receiving the pulse, the slave unit runs for one cycle and stops. If the feedback connection is
made from the SYNC TRIG OUT of the slave unit to the TRIG IN of the master unit, then you can cre-
ate the configuration which indefinetely outputs the signal from each unit upon completion of the other
unit. m this case, you must start the master unit by pressing the manual TRIG button (SHIFT-0) on the
front panel.
5-5
Summary of Contents for 2411B
Page 10: ...1 Quick Start...
Page 22: ......
Page 23: ...2 About Arbitrary Waveforms...
Page 30: ......
Page 31: ...Chapter 2 About Arbitrary Waveforms 2411B Block Diagram 2 8 Block Diagram...
Page 32: ...3 Front Panel Menu Operation...
Page 41: ...4 Waveform Creation System...
Page 43: ...Chapter 4 Waveform Creation System User Interface for the Waveform Creation System 4 3...
Page 61: ...5 Multi Instrument Operation...
Page 68: ...6 Performance Verification...
Page 71: ...Chapter 6 Performance Verification Verification Test Sheet 6 4...
Page 72: ...7 Remote Interface Operation...
Page 78: ...Chapter 7 Remote Interface Operation Command Set Command Set 7 7...
Page 79: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 8...
Page 80: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 9...
Page 81: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 10...
Page 82: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 11...
Page 83: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 12...
Page 84: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 13...
Page 91: ...8 Advanced Remote Operation...
Page 125: ...Appendix...
Page 133: ...AMPLITUDE OFFSET RANGES...
Page 134: ...INDEX...
Page 140: ......