Chapter 1 Quick Start
To set the sample clock rate
To set the sample clock rate.
At the initial power-on, the generator outputs a free-running sine wave at 10kHz with an amplitude of
5.0 V peak-to-peak (into 50
Ω
termination) after the output is turned on. The following steps will show
you how to change the sample clock rate of a user-defined waveform or an optional sequence. The out-
put frequency of any waveform is:
Waveform Frequency = [(Sample Clock) / (Waveform Length)]*(Number of Cycles in the Waveform Frame)
1. Press FUNC (function) key and select a user-defined waveform, WAV#. Select WAV#0 and press
ENTER.
The default length of a user-defined waveform is 2000 points.
2. Press SCLK/FREQ (Sample Clock/Frequency) button.
The displayed sample clock is either the power-on value or the previous sample clock rate selected.
3. Enter the magnitude of the desired sample clock rate.
Use the right-arrow button to place the cursor under the selected digit. Turn the knob to set the digit to
2.
4. Select the range of the desired frequency.
Use the right-arrow button to place the cursor under the multiplier (M). Turn the knob counterclockwise
to switch the range by a decade. Observe the output frequency has changed by a decade.
1-4
SCLK= 10.00 MHZ
FREQ= 5.000 kHz
SCLK= 12.00 MHZ
FREQ= 6.000 kHz
SCLK= 1.200 MHZ
FREQ= 600.0 Hz
Summary of Contents for 2411B
Page 10: ...1 Quick Start...
Page 22: ......
Page 23: ...2 About Arbitrary Waveforms...
Page 30: ......
Page 31: ...Chapter 2 About Arbitrary Waveforms 2411B Block Diagram 2 8 Block Diagram...
Page 32: ...3 Front Panel Menu Operation...
Page 41: ...4 Waveform Creation System...
Page 43: ...Chapter 4 Waveform Creation System User Interface for the Waveform Creation System 4 3...
Page 61: ...5 Multi Instrument Operation...
Page 68: ...6 Performance Verification...
Page 71: ...Chapter 6 Performance Verification Verification Test Sheet 6 4...
Page 72: ...7 Remote Interface Operation...
Page 78: ...Chapter 7 Remote Interface Operation Command Set Command Set 7 7...
Page 79: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 8...
Page 80: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 9...
Page 81: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 10...
Page 82: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 11...
Page 83: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 12...
Page 84: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 13...
Page 91: ...8 Advanced Remote Operation...
Page 125: ...Appendix...
Page 133: ...AMPLITUDE OFFSET RANGES...
Page 134: ...INDEX...
Page 140: ......