Chapter 6 Performance Verification
Introduction
Introduction
This chapter provides a procedure to verify that the Model 2411B has met the specifications published in
the manual within the tolerances.
There are no user accessible calibration points in the unit
. TEGAM,
Inc. recommends an annual calibration verification interval. It is recommended that Chapter 1 Quick
Start should be followed first in order to verify proper operation of the unit. Copy the Verification Test
Sheet at the end of this chapter to record the data.
Test Equipment Required
Analog Oscilloscope:
100 MHz or greater bandwidth.
Frequency Counter:
1 PPM or better accuracy.
True-RMS Digital Multimeter:
5 1/2 digit, 0.1% or better accuracy.
Distortion Analyzer:
0.002% or better THD measurement at 80 kHz bandwidth.
Precision Load
: 50
Ω
, ± 0.1%
Verification Test
Frequency Accuracy
1. Connect frequency counter to 10 MHz Reference Output on the rear panel.
2. Verify that the frequency is 10 MHz ± 50 PPM.
The following procedures require the instrument to be reset to the default settings. Please save the
pertinent waveform file data in the 2411B in the PC by using WaveWorks Pro or other software before
proceeding to the next test.
Reset the 2411B
Turn on the
LINE
(power) switch while holding the
RESET
(Number 9) button down.
Caution:
Reset erases waveform memory and inserts the default files. Store all waveform required for
future use as noted in the preceding step.
Waveform Rise/Fall Time
1. Set Amplitude to 10 V.
2. Select
STDW
(Standard Wave) as
SQU
(Square wave).
3. Connect OUTPUT to an oscilloscope input terminating with a 50
Ω
load.
4. Verify the Rise and Fall times to be less than 150 ns.
6-2
Summary of Contents for 2411B
Page 10: ...1 Quick Start...
Page 22: ......
Page 23: ...2 About Arbitrary Waveforms...
Page 30: ......
Page 31: ...Chapter 2 About Arbitrary Waveforms 2411B Block Diagram 2 8 Block Diagram...
Page 32: ...3 Front Panel Menu Operation...
Page 41: ...4 Waveform Creation System...
Page 43: ...Chapter 4 Waveform Creation System User Interface for the Waveform Creation System 4 3...
Page 61: ...5 Multi Instrument Operation...
Page 68: ...6 Performance Verification...
Page 71: ...Chapter 6 Performance Verification Verification Test Sheet 6 4...
Page 72: ...7 Remote Interface Operation...
Page 78: ...Chapter 7 Remote Interface Operation Command Set Command Set 7 7...
Page 79: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 8...
Page 80: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 9...
Page 81: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 10...
Page 82: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 11...
Page 83: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 12...
Page 84: ...Chapter 7 Remote Interface Operation Command Set continued Command Set continued 7 13...
Page 91: ...8 Advanced Remote Operation...
Page 125: ...Appendix...
Page 133: ...AMPLITUDE OFFSET RANGES...
Page 134: ...INDEX...
Page 140: ......