DS Series Instruction Manual 2006.12
Page 19
6.1.3 Input Mode Setting
You can set the input mode of each channel.
1.
Move the cursor to [MODE] by pressing the down keys at the <INPUT AMP> menu.
2.
Press the Enter key to go to <MODE>.
3.
Move the cursor to the channel to set its input range by pressing the up and down keys. You
can set all effective channels with the same input range at [ALL].
Press the left and right keys to select the desired input mode.
DC
DC coupling mode
IEPE
Turn on the power supply to the sensor connected to the channel. Voltage
output type accelerometer and microphone can be directly connected.
(IEPE = Integrated Electronics Piezoelectric)
AC
AC coupling mode
4.
After setting the input modes of all channels, press the ENTER key to decide the selections.
The <INPUT AMP> menu is displayed.
If you do not want to decide the selection, press the STOP key. The selection is not effective.
Then return to the <INPUT AMP> menu.