CHAPTER 7
Quality control
Sysmex XE-5000 Instructions for Use
7-3
R
e
vi
se
d A
p
ri
l 20
07
X-barM control setting
1. From the Main Menu screen on the Main Unit, select
“QC” on the function menu.
The QC Menu screen will appear.
2. From the QC Menu screen, select “Xm STT/STP” on the
function menu.
• If Xm is displayed in the upper right of the LCD screen,
it indicates that X
M
control is ON.
• If Xm is not displayed in the upper right of the LCD
screen, it indicates that X
M
control is OFF.
Settings for X
M
control can be changed as described below.
1. Start quality control from the Information Processing Unit
(IPU).
2. Click the “Setting” button.
The QC Setting dialog box will be displayed.
3. Click the “X-barM” tab.
The X-barM Setting screen will be displayed.
Information
Even if X-barM control is canceled in this menu when
“Execute X-barM control” is set in the IPU, X-barM
control is executed automatically after the power is
turned on again. Use the menu only when you analyze
a sample in which control error may occur.
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...