CHAPTER 6
Operation
6-18
Sysmex XE-5000 Instructions for Use
Revised July
2
007
8. Mix the sample tube thoroughly by inverting the tube.
9. Remove the cap carefully in order not to splash the blood.
10. Hold the sample tube under the manual aspiration pipette
so that the tip of the pipette is at the bottom of the sample
tube; then press the START switch.
Do not remove the sample tube while the READY LED is
blinking; the sample is being aspirated.
11. After the READY LED turns off (and a short beep sounds
two times), remove the sample tube.
12. When the READY LED turns on again, prepare the next
sample and repeat steps 8 - 11.
Caution!
To remove the sample tube, pull straight down to
prevent bending the manual aspiration pipette.
Otherwise the blood may scatter.
Note:
The manual aspiration pipette is automatically rinsed,
and does not need to be wiped manually.
Information
If a message calling for reagent replacement is
displayed during analysis, replace with new reagent. If
the replace reagent sequence is executed when
reagent amount is insufficient, bubbles may occur,
resulting in higher background value.
READY LED
START Switch
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...