
CHAPTER 10
Troubleshooting
Sysmex XE-5000 Instructions for Use
10-33
R
e
vi
se
d A
p
ri
l 20
07
4. Pump
This program displays the operation cycle count of each
reagent pump.
After replacing the reagents, reset the operation cycle counter
by the procedure described following.
1. From the Main Menu screen on the Main Unit, select
“Test” on the function menu.
The Test Menu screen will appear.
2. Select “Status” on the function menu of the Test Menu
screen.
The Status screen will appear .
3. Select “Pump” on the function menu of the Status
screen.
The Pump screen will appear.
The following operation cycle is displayed on the Pump
screen.
FBA: Indicates the operation cycle count of the FBA
reagent pump.
FFD: Indicates the operation cycle count of the FFD
reagent pump.
FFS: Indicates the operation cycle count of the FFS
reagent pump.
SNR: Indicates the operation cycle count of the SNR
reagent pump.
RED: Indicates the operation cycle count of the RED
reagent pump.
SLS: Indicates the operation cycle count of the SLS
reagent pump.
SIM: Indicates the operation cycle count of the SIM
reagent pump.
4. Select the reagent pump by moving the cursor with
or
key on the Pump screen.
5. When you select “OK” on the function menu of the Pump
screen, then select “OK” again, the highlighted
parameter counter will be reset.
When you wish to cancel the reset operation, select
“Cancel” on the function menu and return to the Status
Menu screen.
Sampler
1. Rack Feed In
Select this program to execute the Rack Feed In operation
test. Set the rack in place before executing the test.
1. Set the rack in place on the right rack pool.
2. From the Main Menu screen on the Main Unit, select
“TEST” on the function menu.
The Test Menu screen will appear.
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
N u m
D P
X m
F B A 5
F F D 5
F F S 3
S N R 6
R E D 4
S L S 2
S I M 8
C D N R
O K
C a n c e l
N o t R e a d y
S e l e c t e d c o u n t e r w i l l b e c l e a r e d . O K ?
< P u m p C o u n t e r >
M a n u a l
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...