
CHAPTER 7
Quality control
Sysmex XE-5000 Instructions for Use
7-9
R
e
vi
se
d A
p
ri
l 20
07
QC data range selection function
The QC data range selection function can be used to perform
the following operations from the QC chart screen in the IPU.
• Select a range and display the Mean, SD, and CV.
• Select a range and use the data in the range to set
automatic targets and automatic limits.
• Select a range and delete the data within it.
• Select a range and output the contents to the GP, LP or host
computer.
Range selection procedure
Use the operations listed below to select a range on the QC
chart screen.
Deleting QC data
If there appears to be a problem with the QC analysis results,
or if incorrect control blood was used, the control data can be
deleted.
• To delete all control data selected with the QC data range
selection function, click “Delete Data”.
• To cancel the delete, click “Undelete”.
Operation
Result
[
←
] key
Moves the cursor one plot to the left.
[
→
] key
Moves the cursor one plot to the right.
[Shift] + [
←
] key
Expands the selected range to the left.
The cursor does not move.
[Shift] + [
→
] key
Expands the selected range to the right.
The cursor does not move.
Mouse left click
Moves the cursor to the mouse pointer
location.
If an area was selected, the selection
changes to a single point.
Shift + Mouse left click
Expands the selected range to the mouse
pointer location.
The cursor does not move.
Drag
Selects a range.
[Ctrl] + [A] key
Selects all data for the range.
The cursor moves to the position of the
newest data.
Information
Plots that have been deleted will disappear from the
display, but they are retained in internal data until the
tab is changed or the QC chart screen is closed, so
undo is possible.
Once the tab has been changed or the QC chart has
been closed, the plots can no longer be undeleted.
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...