CHAPTER 8
Calibration
8-8
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
Updating the calibration values
1. After entering the calibration value, select “Execute” on
the function menu of the Manual Calibration screen.
2. Both the old and updated calibration values of HGB and
HCT will appear on the Manual Calibration Setting screen.
3. Select “Continue”, “OK” or “Cancel” on the function
menu.
Continue: Returns the system to the Manual Calibration
screen and the operation can be continued.
OK:
Sets the updated calibration value and
returns the system to the Calibration Menu
screen.
Cancel:
Cancels the updated calibration value and
returns the system to the Manual Calibration
screen.
When the Calibration Error message is displayed, select
“OK” on the function menu to return the system to the
Manual Calibration screen. The calibration can be
continued.
4. Reanalyze the sample used for the calibration using the
XE-5000. Confirm that the analysis value is within the
allowable range and does not vary greatly from the
reference value.
Recalibrate if HGB and HCT values are consistently
higher or lower overall than the reference value. If, after
re-calibration, the analysis values are still outside the
allowable range, or if abnormal data is found, check the
samples for abnormalities such as abnormal blood
coagulation, blood cell morphology, patient medicinal use,
and old blood. If no abnormality is found on the samples,
contact your Sysmex technical representative.
Note:
• If you press the RETURN key, the updated
calibration value will be canceled, and the system
will return to the Calibration Menu screen.
• When the updated calibration value exceeds
100±20%, the Calibration Error message will be
displayed and the Manual Calibration Setting screen
will not appear.
• When a calibration over ±20% is necessary, contact
your Sysmex technical representative.
C o n t i n u e
O K
C a n c e l
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
M a n u a l
H G B
H C T
1 0 0 . 0 %
→
1 0 0 . 7 %
1 0 0 . 2 %
→
1 0 0 . 5 %
C
N o t R e a d y
< M a n u a l C a l . >
N u m
D P
X m
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...