Executing the manual calibration program
1. From the Main Menu screen on the Main Unit, select
“Calibration” on the function menu.
The Calibration Menu screen will appear.
2. Select “Manual Cal.” on the function menu of the
Calibration Menu screen.
The Manual Calibration screen will appear.
Entering the calibration values
1. Using the
and
keys, move the cursor to select HGB
or HCT.
2. Using the numeric keys, enter the calibration value
obtained by calculation.
3. Press the ENTER key.
The entered value is set and the cursor will move to the
next parameter.
Note:
When the “Calibration” on the function menu is not
displayed on the Main Menu screen, press the MORE
key to change over the function menu.
E x e c u t e
N e x t N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 6
D P N o . 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5
M a n u a l
H G B
H C T
1 0 0 . 7 %
1 0 0 . 0 %
C
N o t R e a d y
< M a n u a l C a l . >
N u m
D P
X m
Note:
Input calibration values in the range 80.0 – 120.0 to
the first decimal place.
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...